UP-7000 Multi-parameter
Features
• 12.1" high-resolution TFT display with LED backlight
• Arrhythmia analysis and S-T segment measurement
• Protection against defibrillator discharge
• Adult/Pediatric/Neonate measurement modes
• Visual and audible alarms; Networking capability
• Up to 9 waveforms simultaneously display
• 140-hour ECG waveform data storage and recall
• 2000-hour data trends with a graphic and tabular view
• 2000 groups event, ARR and SpO2 storage
• Built-in lithium battery; Touch screen optional
Configurations
NIBP | Technique | Oscillometric |
Typical measuring time | <30 seconds (typical adult cuff) | |
Initial cuff inflation pressure |
Adult: <175mmHg | |
Pediatric: <135mmHg | ||
Neonate: <65mmHg | ||
Overpressure protection limit |
Adult: 300mmHg | |
Pediatric: 240mmHg | ||
Neonate: 150mmHg | ||
Measuring range | ||
Systolic pressure | Adult: 40mmHg~275mmHg | |
Pediatric: 40mmHg~200mmHg | ||
Neonate: 40mmHg~135mmHg | ||
Diastolic pressure |
Adult: 10mmHg~210mmHg | |
Pediatric: 10mmHg~150mmHg | ||
Neonate: 10mmHg~95mmHg | ||
Mean arterial pressure |
Adult: 20mmHg~230mmHg | |
Pediatric: 20mmHg~165mmHg | ||
Neonate: 20mmHg~110mmHg | ||
Measurement accuracy |
Maximum mean difference: ±5 mmHg | |
Maximum standard deviation: 8 mmHg | ||
Measurement mode | Manual, Auto, STAT | |
Automatic measuring intervals | 1~480min | |
TEMP | Measuring range | 21.0°C~50.0°C |
Measuring accuracy | ±0.2°C for range from 25.0°C~45.0°C | |
SpO2 | Transducer | Dual-wave length LED |
SpO2 measuring range | 0%~100% | |
SpO2 measuring accuracy | 2% for range from 70% to 100% | |
Low perfusion performance | As low as 0.4% | |
PR measuring range | 0bpm~250bpm | |
PR measuring accuracy | ±2bpm or ±2%, whichever is greater | |
ECG | Input dynamic range | ±0.5mVp~±5mVp |
HR measuring range | 15bpm~350bpm | |
HR measuring accuracy | ±1% or ±2bpm, whichever is greater | |
HR alarm delay time | ≤10s | |
Sensitivity selection | ×1/4, ×1/2, ×1, ×2, ×4 and Auto | |
Sweeping speed | 6.25mm/s,12.5mm/s, 25mm/s, 50mm/s | |
ECG noise level | ≤30µVp-p | |
ECG input loop current | ≤0.1µA | |
Differential input impedance | ≥10Mohm | |
Common-mode rejection ratio(CMRR) |
≥105dB (Monitoring mode) | |
≥89dB (Diagnostic mode) | ||
Time constant |
≥0.3s (Monitoring mode) | |
≥3.2s (Diagnostic mode) | ||
RESP | RR measuring range | 0rpm~120rpm |
R measuring accuracy | ±5% or ±2rpm, whichever is greater | |
Others | Power supply | 100~240Vac, 50/60Hz |
Built-in battery | 4400mAh Lithium battery | |
Display | 12.1 inch (Resolution 800*600) | |
Alarming mode | Audible-visual alarm | |
Networking port | Ethernet port | |
Standard Configuration | ECG, RESP, SpO 2, NIBP, TEMP, PR | |
Option | 2-IBP, EtCO, Nellcor Spo, SunTech NIBP, Cardiac Output, Built-in printer, Cerebral State Monitoring, Central monitoring system, Touch screen |